OI l-L ess Compressor Installation A130110m
by MIDMAIRK
For models:
021. 022, 062, 082
Step 1: Remove compressor and accessories
from shipping skid.
7_ -
Step 3: Move compressor to a dry, well ventilated area on a solid, level surface.
Note: Verify compressor is level by observing bubble level on saddle.
Temperature of room 40°(4°C)min. - 100°(38°)max.
J
7/
41111t/N;',
Bubble � �
Level ;� ;f;
' 11 40' *(��11r ),..,,„,,,,,fr
t'!SIN': 011 4&. *,
--b-wor.4 '611111 11,-,?,47177---:-, .
1 Drain Canister i
• 1 0 40- 04.14;!. /7-;)) •P ;41
---..._-------,.%/ 40-"Ill L:10 0 ; '7-t:
Step 5: Install canister 11110 ► ' 062 Model Shown
tubing.
10 I -, AA 148900i ,-
CStep 2: Install feet) IS .�°
Step 4: Connect air line to office air supply line.)
1 Step 6: Mount fresh air intake manifold.
Connect to air source and compressor intake mufflers.
Note: Mount no more than 3'above compressor suction intake(s).J
i
Equipment Alert
ATight bends will cause hose to collapse and lead to compressor failure.
To prevent moisture from entering the compressor. remote fresh air intake air temperature must not exceed
compressor room air temperature by more than 20°. If necessary, route the intake to a source inside the
building away from the compressor room.
Single Head Compressors-021&022 Twin Head Compressors-062&082
To Acceptable Air Intake Source To Acceptable Air Intake Source
/1!1/7 PVC Pipe I2"PVC Pipe
Wall Bracket 2 x 12 Bushing Wall Bracket
1 1/7•'x 12" ' y' ,�
Bushing ii I' X112"Tee •
12"Hose Clamp
1/2"Hose 2"Cross
1/2"Hose 12"Hose Clamp■
Barb Adapter 1 �, Barb Adapter 41,,,
iii 1/2"Nylon Braid 1/2"Nylon Braid
Hose Hose
(5'Max Length) rap (5'Max Length)
2•'
Tee
To Each Muffler on Compressor Head To Muffler on Each Compressor Head
AA 149400
Fresh Air Intake -9.-- F:::1)......; ...,,
4011411 I 4 I 6' l
;fr 1 G,1 Ill\
t„,„,„ ,,,,,,...
, ,._:_____,* „,,.
6 .,„. •
AA 1492001
062 Model Shown
Specification Sheet olio
Oil-Less Compressors
R Apw
by MIDMARK
Itt PVC Air intake Vent
(Provided By Plumber)
1
'
,�'r/ witDisconnect isted Box Wall Mounted 24 V Compressor
'ii�l ,i ��,\ Breaker/Fuse Starter Contacor(PN ACACSC)
�7��PP as per Specification
( 4,,N, 111:::: 11,74,
kti4i- 4 ofv.t7-, cd 44. 0 e
el
_ _ . 1 sl•wr '0
y�i ill!� '
1
Condenstae Drain Canister ��� ii,,,,,...„ �V =(Provided with 5'of 3B" . 0 +i
Polyflo Tubing) • 1
Level Compressor with "- r `` ` Air Line(Provided by Plumber)
Isolation Feet(Provided) °\
/ 5'Flex Hose with 12"Or 3/8"MPTTermination(Provided)
IAA 14600011
Breaker
Vacuum Max. Width Depth Height Weight Tank Total Amps Size per CFM @ 80 Sound
Model Users Heads (IN.) (IN.) (IN.) (LBS.) Capacity HP Voltage Per Pump PSI Level
Pump (Amps) (DB-A)
021 3 1 24 21 29 208 10 1.5 115 20 30 5.2 78
062 7 2 38 31 33 410 30 3 208 -230 20 30 12.4 81
082 10 2 38 31 33 497 30 4 208-230 20 30 16.9 82
\. f